|
||||||||||||
|
||||||||||||
|
||||||||||||
|
||||||||||||
|
||||||||||||
|
||||||||||||
AIMS of STUDY
- good urodynamics (UDS) depends on the accurate measurement of pressure during
filling and voiding, normally using either a fluid-filled catheter connected
to an external transducer (ET) or by a catheter-mounted microtransducer (MT).
ETs are prone to problems such as air bubbles, leaks or kinks in connecting
tubing, and damping of pressure transmission. MTs are delicate and temperature-sensitive.
Once inserted they cannot be re-zeroed to atmospheric pressure without being
removed. Artefactual pressure changes may occur due to impact of the transducer
with faeces or with movement. An important theoretical objection to the use
of MTs relates to the difference in the pressure measured by the different systems1.The
pressure recorded by an ET depends on the level of the ET in relation to the
body. If the ETs measuring abdominal and vesical pressure (pabd and pves) are
at the same level, in the absence of active contraction or passive distension
the recorded pressures should be the same, and thus detrusor pressure (pdet)
will normally be close to zero. On the other hand, the pressure measured by
an MT depends on its position in the body cavity, which is not normally known.
As rectal and vesical MTs may well be at different levels in the body, the pdet
will tend to be inaccurate. Absolute pressures measured by MTs will also differ
from those recorded by ETs. We were unable to identify any published data on
the magnitude of the difference in recorded pressures at rest, or on the measurement
of other pressure changes (e.g. due to detrusor contraction or coughing) by
the two systems, and therefore designed the present study.
METHODS
- women referred for UDS were recruited. Pressures were recorded throughout
UDS using MTs and ETs, with simultaneous pressure subtraction using a Dantec
Duet machine. Both systems were zeroed to atmospheric pressure, with ETs fixed
at the level of the symphysis pubis. A single transducer Gaeltec MT catheter,
and an epidural catheter piggy-backed on an 8 Fr filling catheter, were inserted
together into the bladder. A single-transducer Gaeltec MT catheter and a 6 Fr
catheter, protected by finger cots, were inserted together into the rectum.
Equal pressure transmission was tested by coughing, and connecting tubing for
ETs was checked for bubbles, leaks or kinks.
The pabd, pves and pdet were recorded at rest in different postures before UDS, which was performed with the patient sitting, filling at 30-50 ml/min and removing the filling catheter at capacity. Computer traces were reviewed by a single reviewer (J.S.) for the measured pressures during coughs, unstable detrusor contractions, voiding contractions, and after-contractions (ACs). Traces were also assessed for the overall quality of cough and live signal subtraction.
RESULTS
- 20 women agreed to join the study. One woman found catheter insertion too
painful, leaving 19 tests for analysis. The mean pressures recorded using ETs
and MTs are shown in the table below. All means are in cmH2O.
|
Posture |
SUPINE |
SITTING |
STANDING |
||||||
|
Pressure Measured |
Pves |
Pabd |
Pdet |
Pves |
Pabd |
Pdet |
Pves |
Pabd |
Pdet |
|
External Transducer
(ET) |
6.2 |
5.1 |
1.1 |
28.9 |
28.4 |
0.5 |
35.8 |
34.7 |
1.0 |
|
Micro Transducer
(MT) |
11.4 |
16.1 |
-4.7 |
32.2 |
35.2 |
-3.1 |
33.0 |
33.4 |
0.4 |
|
Mean Diff.
between ET and MT |
5.2 |
11.0 |
5.8 |
3.3 |
6.8 |
3.6 |
-2.8 |
-1.3 |
1.4 |
Cough subtraction was normally good with both systems. Live signal subtraction was generally more precise with MTs, giving a smooth pdet trace. Detrusor instability was noted in 8 patients. In 3 of these the measured amplitude of detrusor contractions was markedly different (>20 cm H2O). In all 3 cough tests suggested reasonable pressure subtraction, although in 2 of the 3 there were other suggestions of imperfect pressure transmission. Four patients were unable to void with catheters in, and 2 patients had 1 or more lines displaced during voiding. In 3/13 patients with complete voiding data the pdet at maximum flow differed by >20 cm H2O, again with reasonable cough subtraction in all 3, but with some other evidence of imperfect pressure transmission in 2 patients. ACs were noted in 4 of the patients who voided. In all 4 cases, the peak pressure measured by MT during ACs was higher than with ET (+23 to +168 cm H2O), despite otherwise good evidence of equal pressure transmission in 3 out of 4.
CONCLUSIONS
- the differences in absolute resting pressures recorded using ETs and MTs are
relatively modest, but should perhaps be borne in mind when analysing absolute
pabd or pves values, e.g. during leak point pressure measurement. Differences
in resting pdet are small, and in the sitting and standing positions are not
likely to have a signifcant impact on the interpretation of UDS results. However,
comparison of simultaneous measurement with ETs and MTs showed moderate or large
differences in the magnitude of changes in pdet in 7 of 19 patients, despite
apparently good pressure transmission in both systems in most cases. Whether
these differences reflect problems with ETs (e.g. kinking of fluid-filled tubing),
or MTs (e.g. direct contact of the bladder wall with the transducer), or both,
is unclear. This small study suggests that it cannot be assumed that detrusor
pressure changes are measured equally by different measurement systems. The
quantification of pdet changes with ETs and / or MTs - e.g. in diagnosing bladder
outlet obstruction or in grading detrusor instability - may be less accurate
than previously thought. As both systems are widely used, this is an area worthy
of further study.
REFERENCES
- 1. Proceedings of the 12th Annual Meeting of the ICS, Leiden, 1982, pp 53-5.
ACKNOWLEDGEMENTS
- we would like to thank Lorex Synthélabo and Southmead Hospital Research Foundation
for their financial support. We would also like to thank Adrian Smart of Dantec
UK for his advice and for the loan of the Dantec Duet machine used in the study